What an inspector reads
Inspectors read what reports leave out.
Today
- Dynamic conditions not described
- Hoods drop out between visits
- A pass with no as-found value
With Laminar
- Every gap on one page
- Hood by hood, visit by visit
- A list to hand your certifier
In the Laminar record
Every hood, on its six-month clock.
Each hood and room carries its due date and the rule behind it.
- Hood 1BSC9 days
- Hood 2Isolator (CACI)4 mo
- BufferISO 72 mo
- AnteISO 85 mo
503B outsourcing facility?
The gap check also reads for CGMP: in-operation classification and the as-found and retest trail.
Fair questions
Before you change a workflow.
Criteria built from
- ISO 14644-1
- ISO 14644-3
- USP <797>
- USP <800>
- EU GMP Annex 1
- IEST-RP-CC034
What comes back?
One page, by email: what dropped out, what moved and what an inspector is likely to ask. It compares what your reports contain; it is not a rating of your certifier. We sign your NDA before you send a single report. Your reports are read only by the person preparing your summary, and deleted within 30 days of the check. The summary comes back within 5 business days. Privacy notice.
What does it cost, and who pays?
The report gap check is free. Your first job on Laminar is free, with no commitment. After that, Laminar is priced per room or per visit, quoted for your rooms on the call. No setup fee: we build your facility from your drawings and past reports. Your pharmacy can hold the account and add your certifier's technicians to it, so the record of every hood and room stays with you, whichever certifier comes next.
Does Laminar certify our rooms?
No. A qualified certifier tests with calibrated instruments and signs. Laminar is the software the results are recorded in, so every result behind that signature keeps its as-found value, adjustment, retest, instrument and time.
We're a 503B outsourcing facility. Does this fit?
Yes. Outsourcing facilities follow FDA CGMP requirements as well as USP <797>. For a 503B, the gap check also looks for in-operation classification and the as-found and retest trail across your reports. In Laminar: ISO 14644-1 classification at rest and in operation, HEPA leak tests, pressure differentials in Pa, the as-found and retest trail, review records and the audit trail. See how QA and GMP sites use it.
More questions (6)
What does our pharmacy get if our certifier issues reports in Laminar?
When your certifier issues a report in Laminar, they can give you secure sign-in access to it: every result, every as-found value, every retest and the evidence behind it, exactly as issued. Drafts stay with your certifier. A revision is a new snapshot, and nothing is ever overwritten. A full facility portal, with every room's history behind one login, is coming next. You don't need to test anything yourselves.
What should a complete certification report show?
- PEC airflow and smoke studies, with the dynamic conditions described
- HEPA filter integrity for each hood and room
- Total particle counts, at rest and in operation
- Air changes and room pressure differentials, buffer and ante room
- As-found and retest for anything adjusted
- Instrument and calibration on every test
- USP <800> hazardous-drug rooms: negative pressure of at least 0.01 in. w.c. (2.5 Pa) recorded as a value, at least 12 air changes per hour, and the C-PEC externally vented
What if our certifier won't use Laminar?
Then nothing changes for you. Your certifier keeps the job, and the gap check still tells you exactly what to ask for on your next report. If you'd like them to, here is a note to send:
Will this help at a board of pharmacy, NABP or FDA inspection?
Yes. The summary lists what an inspector is likely to ask about, written so you can hand the list straight to your certifier and get corrected or complete reports before the inspection date. Laminar never certifies your rooms and never rates your certifier: it shows what the reports contain.
Which hoods does Laminar cover?
Each hood carries its own test package: biosafety cabinets built around NSF/ANSI 49 methods (downflow, inflow, HEPA leak scans, work-zone counts, smoke and alarms), clean benches, CAI/CACI and isolators. NSF/ANSI 49 values are marked to verify against a licensed copy before a report relies on them.
Inches of water column or Pa?
Your choice, set per job. USP <797>'s minimum of 0.02 in. w.c. is about 5 Pa.
Free report gap check
Get your free report gap check.
Name, email and your pharmacy. We reply with the NDA and how to send your reports.
Book a 30-minute callor send the form and we reply within one business day.
David WeilFounder · a named person, not a ticket queue
- 1 business dayReply, guaranteed
- NDA firstBefore any report
- First job freeNo commitment
Gap check back in 5 business days · Reports deleted within 30 days · Originals never overwritten
Got it — thank you.
We reply within one business day with the NDA and how to send your reports. Hold the reports until then. Emails come from marketing@laminarcert.com; add it to your contacts so it doesn't land in spam.
Urgent? Call +1 (704) 500-8100 and mention your facility.